Good price  online

products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Radon Detector
Created with Pixso. RAD-S51 Radon Gas Sensor Silicon Detector 5-300000 Bq/m3 Continuous Indoor Air Radon Monitor

RAD-S51 Radon Gas Sensor Silicon Detector 5-300000 Bq/m3 Continuous Indoor Air Radon Monitor

Brand Name: CESTSEN
Model Number: RAD-S51
MOQ: 1 Piece
Delivery Time: 3-7 working days for stock, 15-25 days for production
Payment Terms: T/T
Detail Information
Place of Origin:
Shenzhen, Guangdong, China
Certification:
CE, RoHS
Detector Type:
Silicon Detector
Measurement Object:
Radon Gas
Measurement Range:
5 ~ 300000 Bq/m3
Detection Sensitive Area:
100 Mm2 (10 X 10 Mm)
Sensitivity:
0.027 Cph/[Bq/m3] Or 0.982 Cph/[pCi/L]
Detection Limit:
5 Bq/m3
Background Count Rate:
0.05 Cph (Nitrogen) Or Less
Uncertainty:
120 Hours Later 10% (K=2) When Radon Concentration Greater Than 200 Bq/m3
Measurement Time:
60 Min Or More
Working Voltage:
3.3V
Power Consumption:
3.3 MW Or Less
Communication Protocol:
3.3V TTL Level Serial Asynchronous Output
Baud Rate:
9600 Bps
Working Temperature:
-10 To +50 DegC
Working Humidity:
90% Or Less (at +45 DegC)
Storage Temperature:
-25 To +85 DegC
Dimensions:
60 X 60 X 21 Mm
Packaging Details:
CESTSEN branded color box with shock-proof foam insert, export-grade corrugated carton
Supply Ability:
3000 Pieces per Month
Highlight:

RAD-S51 radon gas sensor

,

continuous indoor air radon monitor

,

radon detector with silicon sensor

Product Description

RAD-S51 Radon Gas Sensor

The RAD-S51 radon detector mainly measures the concentration of radon gas in the environment through static diffusion of air. By utilizing the radioactive properties of radon gas and its decay products, the detector obtains alpha counts through a silicon detector and a signal acquisition system, and calculates the radon gas concentration to achieve long-term continuous monitoring of radon levels in the environment.

RAD-S51 Radon Gas Sensor Silicon Detector 5-300000 Bq/m3 Continuous Indoor Air Radon Monitor

Product Features

  • The measurement system is not sensitive to humidity.
  • With a large-sized collection chamber, the measurement sensitivity is high.
  • It is used for measuring the radon concentration in indoor environments.
  • The parameters can be adjusted to meet various data requirements.

Application Fields

  • Occupational health environment radon gas measurement
  • Static radon measurement in industrial and mining environments
  • Regional long-term air radon monitoring in ecological environment

Technical Indicators

Indicators Parameters
Detector Silicon detector
Measurement object Radon gas
Measurement range 5 ~ 300000 Bq/m3
Detection sensitive area 100 mm2 (Size 10 × 10 mm)
Sensitivity 0.027 cph/[Bq/m3] or 0.982 cph/[pCi/L]
Detection limit 5 Bq/m3
Background count rate ≤ 0.05 cph (Nitrogen)
Uncertainty 120 hours later ≤ 10% (K=2) (when radon concentration is greater than 200 Bq/m3)
Measurement time ≥ 60 min
Working voltage 3.3V
Power consumption ≤ 3.3 mW
Communication protocol 3.3V TTL level serial asynchronous output
Baud rate 9600 bps
Work environment Temperature: -10°C ~ +50°C; Relative humidity: ≤ 90% (at +45°C)
Storage environment Temperature: -25°C ~ +85°C; Relative humidity: ≤ 90% (at +45°C)
Dimensions 60 × 60 × 21 mm